Creatine Kinase Isoenzyme MB (CK-MB) Test Kit (Fluorescence Immunochromatography)
Share:

Creatine Kinase Isoenzyme MB (CK-MB) Test Kit

(Fluorescence Immunochromatography)


[Product name]

Creatine Kinase Isoenzyme MB (CK-MB) Test Kit (Fluorescence Immunochromatography)

[Packaging specifications]

10 tests/box, 20 tests/box, 25 tests/box, 30 tests/box, 50 tests/box, 100 tests/box.

[Intended use]

This kit is used for the quantitative determination of Creatine Kinase Isoenzyme MB (CK-MB) in human serum, plasma or whole blood samples in vitro.

There are four isozyme forms of Creatine Kinase (CK): muscle type (MM), brain type (BB), mixed type (MB) and mitochondrial type (MiMi), among which MB type is mainly present in myocardial cells. In myocardial infarction, creatine kinase increases within 6 hours of onset, peaks at 24 hours, and returns to normal within 3~4 days. Among them, Creatine Kinase Isoenzyme MB MB has high diagnostic specificity, so it has become one of the markers of myocardial infarction.

[Test principle]

This test kit uses fluorescence immunochromatography technology to quantitatively determine the content of Creatine Kinase Isoenzyme MB (CK-MB) in serum, plasma or whole blood samples by double antibody sandwich method. After the sample is added to the test kit, the CK-MB in the sample reacts with the fluorescently labeled CK-MB monoclonal antibody 1 to form a complex, which continues forward chromatography along the nitrocellulose membrane to the test line, and forms a complex binding with coated CK-MB monoclonal antibody on the test line. Under the action of the excitation light source, the content of Creatine Kinase Isoenzyme MB (CK-MB) in the sample is positively correlated with the fluorescent antibody signal, and the content of Creatine Kinase Isoenzyme MB (CK-MB) in the sample was calculated by fluorescence immunoanalyzer.

Principle of the supporting instrument: The measuring system of the instrument scans the binding area between the marker and the analyte on the test kit after the reaction to obtain the optical signal. Then the optical signal is measured and analyzed to quantitatively obtain the concentration of the analyte.


 [Main components]

No.

Name

Component

1

Creatine Kinase Isoenzyme MB (CK-MB) Test Kit   (Fluorescence Immunochromatography)

Each bag contains a test card and a desiccant; the test   card consists of shell and a test strip, and the test strip consists of   sample pad, fluorescent pad (fixed with fluorescently labeled CK-MB   monoclonal antibody 1), Nitrocellulose membrane (coated with CK-MB monoclonal   antibody 2 and goat anti-mouse IgG), filter paper and PVC plastic sheet

2

Whole Blood Diluent

1 bottle of diluent for 10 tests/box, 20 tests/box, 25   tests/box and 30 tests/box; 2 bottles of diluent for 50 tests/box; 4 bottles   of diluent for 100 test/box; 3.0mL/bottle

3

ID card

1 pc/box

[Applicable instrument]

Fluorescence immunoassay analyzer

[Storage conditions and validity period]

The test kit should be stored in a sealed aluminum foil bag at 4℃30℃, and the validity period is 18 months. After the foil bag is opened, the validity period is 1 hour.

[Sample requirements]

1. Human serum, plasma or whole blood samples; other body fluids and samples may not give accurate results.

2. Venous blood or fingertip blood should be collected under sterile conditions. It is recommended to use human serum or plasma for testing.

3. Anticoagulation with EDTA, sodium citrate or heparin is recommended for plasma and whole blood samples.

4. After the clinical blood samples are collected, the test must be completed within 4 hours at room temperature; serum and plasma can be stored at 2~8°C for 3 days and stored below -20°C for 5 months. Whole blood samples should not be frozen and stored at 2~8°C for 3 days. Avoid heat inactivating samples, and hemolyzed samples should be discarded.

5. Samples must be returned to room temperature before testing. Refrigerated samples need to be completely thawed, rewarmed, and evenly mixed before use. Do not freeze and thaw repeatedly.

[Test method]

Please read the instructions of the test kit and the manual of the fluorescence immunoanalyzer carefully before use.

a)         Bring the test kit and sample to be tested to room temperature.

b)        Make sure the ID card matches the batch number of the kit, and insert the ID card into the card reading area of the instrument to read the information.

c)         Open the inner package of the test kit, take out the test card; draw 70μL of serum plasma sample, drop vertically to the sampling place of the test card, and start timing; draw 70μL whole blood, drop vertically to the sampling place of the test card, and immediately add 1 drop of whole blood diluent at the sampling place and start timing.

d)        After adding the sample, click "Start Test" on the screen of the fluorescence immunoanalyzer, and the test card will react at room temperature for 10 minutes; insert the test card into the test card slot of the fluorescence immunoanalyzer, and the instrument will automatically test the test card; The test results can be seen on the display screen of the fluorescence immunoanalyzer. Click "Print" on the screen to print the results.

[Positive judgment value or reference interval]

Normal reference value: <5ng/mL

Due to differences in geography, race, gender and age, it is recommended that each laboratory establish its own positive judgment value or reference interval.

[Interpretation of test results] (For reference only, not used as clinical diagnostic criteria, test results need to be combined with other clinical and laboratory data for clinical diagnosis)

Content of CK-MB (ng/mL)

Clinical application advice

0-5

Normal level

5

It is suggested that the patient is at risk of   myocardial infarction. Note: CK-MB exceeds the normal level within 4~8 hours   after myocardial infarction; peaks within 12~24 hours; and returns to normal   level in about 3 days.

 

[Quality Control]

Quality control should be carried out on a regular basis to ensure the validity and accuracy of test results. The test kit does not contain any quality control products, it is recommended to select appropriate quality control products for quality management.

[Limitations of the test method]

a)         This test kit is only for the test of human serum, plasma or whole blood samples. Abnormal hematocrit samples have an impact on the results of the whole blood test. The test results of serum, plasma and whole blood are not significant difference when the hematocrit is between 21% and 48%.

b)        The test results need to be combined with other clinical and laboratory data, and if the CK-MB test results do not match the clinical assessment, further testing is required.

c)         False positive results may be caused by: cross-reaction of antibody-like components in the blood; some nonspecific components in blood have similar antigenic determinants to capture fluorescent-labeled antibodies.

d)        Hemoglobin, triglycerides and bilirubin in the blood sample can interfere with the test results, with the maximum allowable concentrations being 5g/L, 10g/L and 0.2g/L, respectively.

e)         False negative results may be caused by: some unknown components shield the antigenic determinants from binding to the antibody; the unstable CK-MB antigen gradually degrades with time and temperature and cannot be recognized by the antibody. Effective test results depend on good reagent and sample storage environment.

f)         Other factors, including technical reasons, operational errors and other sample factors, may also cause errors in test results.

[Product performance index]

a)         Accuracy: The recovery rate is between 85% and 115%.

b)        Linear range: within the linear range of 2.5ng/mL~80ng/mL, the linear correlation coefficient r 0.9900;

c)         Blank limit: not higher than 2.5ng/mL;

d)        Precision:

In-batch precision: The coefficient of variation (CV) is not more than 15%;

Precision between batches: The relative range between batches is not more than 15%.

[Precautions]

1)        This product is a single-use in vitro diagnostic reagent. Do not use expired products.

2)        The ID card and the test card cannot be used together, if they are different batches.

3)        Please read the instructions of the test kit and the manual of the fluorescence immunoanalyzer carefully before  use.

4)        The environmental temperature should be 15~30 when testing. The test card stored at low temperature needs to be returned to room temperature before opening to avoid moisture absorption.

5)        The test card and fluorescence immunoanalyzer should avoid vibration and electromagnetic environment when using. It is normal for the instrument itself to vibrate during normal use.

6)        There is a desiccant in the aluminum foil bag, do not take it orally.

7)        Appropriate biosafety assurance procedures should be in place for those substances containing or suspected of containing infectious agents. The following are the relevant precautions: Handle the sample with gloves or disinfect this reagent; disinfect spilled samples or reagents with disinfectant; disinfect and handle all specimens, reagents and potential contaminants in accordance with local regulations.

8)        This product is used for rapid clinical diagnosis and screening, and the final diagnosis should be made by a doctor after comprehensive examination indicators and clinical symptoms.

9)        The used material must be discarded immediately after test. Disposal should be in accordance with accepted medical practice and applicable national, local or institutional guidelines.  

10)      Do not use if package is open or damaged.  